• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY INTERNATIONAL LTD - 8010379 9/10 COCR HEAD 28MM +3; SMALL TAPER (9/10) FEMORAL HEADS : HIP METAL FEMORAL HEADS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY INTERNATIONAL LTD - 8010379 9/10 COCR HEAD 28MM +3; SMALL TAPER (9/10) FEMORAL HEADS : HIP METAL FEMORAL HEADS Back to Search Results
Model Number 1521-90-059
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Hypersensitivity/Allergic reaction (1907); Renal Failure (2041); Joint Dislocation (2374); Osteolysis (2377); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 01/01/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
This pc is related to (b)(4).This pc reports about the second dislocation and osteolysis.It was reported that on an unknown date, 1991, the patient underwent the tha surgery with unknown cup and stem.After the surgery, on an unknown date, the loosening of the cup and stem occurred, and the first revision surgery was performed on (b)(6) 2010.After the surgery, the patient dislocated on an unknown date, 2016.The surgeon commented that ct can show widespread tumorous appearance around the hip.Osteolysis can also be confirmed.It may be a pseudotumor symptom due to cobalt chromium.Currently, reoperation is postponed due to decreased renal function.On an unknown date 2021, the patient dislocated again and x-ray showed osteolysis of the pelvis and femur.The second revision surgery will be performed if necessary and the liner and cup will be replaced.No further information is available.
 
Additional Manufacturer Narrative
Product complaint # : (b)(4).Investigation summary : no device associated with this report was received for examination.Visual examination of the provided x-ray images confirmed the reported dislocation.A worldwide lot specific complaint database search, or device history record (dhr) review, was not possible because the required lot number was not provided.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Device history lot : a worldwide lot specific complaint database search, or device history record (dhr) review, was not possible because the required lot number was not provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
9/10 COCR HEAD 28MM +3
Common Device Name
SMALL TAPER (9/10) FEMORAL HEADS : HIP METAL FEMORAL HEADS
Manufacturer (Section D)
DEPUY INTERNATIONAL LTD - 8010379
st. anthony's road
leeds LS11 8DT
UK  LS11 8DT
MDR Report Key12159923
Report Number1818910-2021-15050
Device Sequence Number19705696
Product Code JDI
UDI-Device Identifier10603295276647
UDI-Public10603295276647
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K871867
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2010
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1521-90-059
Device Catalogue Number152190059
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/19/2021
Supplement Date Received by Manufacturer08/26/2021
Initial Report FDA Received Date07/13/2021
Supplement Report FDA Received Date08/29/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
9/10 COCR HEAD 28MM +3; SOLUTION A 8IN BOW 16.5 RT; 9/10 COCR HEAD 28MM +3; SOLUTION A 8IN BOW 16.5 RT
Patient Age64 YR
-
-