• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: A.I.D.D LONGFORD ARCHITECT TOTAL B-HCG REAGENT KIT; SYSTEM, TEST, HUMAN CHORIONIC GONADOTROPIN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

A.I.D.D LONGFORD ARCHITECT TOTAL B-HCG REAGENT KIT; SYSTEM, TEST, HUMAN CHORIONIC GONADOTROPIN Back to Search Results
Model Number 7K78-30
Medical Device Problem Code False Negative Result (1225)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/20/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
There was no further patient information provided by the customer.An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.
 
Event or Problem Description
The customer observed false negative architect total b-hcg result on one patient.The results provided were: on 20jun2021 sid (b)(6) initial =<1.2 miu/ml (or=25.00 miu/ml=positive) there was no reported impact to patient management.
 
Additional Manufacturer Narrative
The complaint evaluation included a review of data and information provided by the customer, search for similar complaints, ticket trending review, labeling review, device history record review, field data and in-house testing of architect total b-hcg reagent lot 21034ui00.Trending review determined no adverse trend for the issue for the product.Return testing was not completed as returns were not available.Device history record review did not identify any non-conformances or deviations with the complaint lot and complaint issue.The overall performance of architect total -hcg reagents in the field was reviewed using data gathered via abbottlink from customers worldwide and suggested that the performance of the lot is acceptable.Inhouse testing determined that the specificity performance is not negatively impacted.A review of the manufacturing documentation did not identify any issues associated with the customer¿s observation.A review of the product labeling concluded that the issue is sufficiently addressed.Based on the investigation no product deficiency was identified for the architect total b-hcg reagent lot 21034ui00.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCHITECT TOTAL B-HCG REAGENT KIT
Common Device Name
SYSTEM, TEST, HUMAN CHORIONIC GONADOTROPIN
Manufacturer (Section D)
A.I.D.D LONGFORD
lisnamuck
co. longford
longford N39E9 32
EI  N39E932
MDR Report Key12160114
Report Number3005094123-2021-00132
Device Sequence Number19978688
Product Code DHA
UDI-Device Identifier00380740014971
UDI-Public00380740014971
Combination Product (Y/N)N
Initial Reporter CountryRS
PMA/510(K) Number
K983424
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/01/2021
Device Model Number7K78-30
Device Catalogue Number07K78-30
Device Lot Number21034UI00
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/25/2021
Supplement Date Received by Manufacturer07/19/2021
Initial Report FDA Received Date07/13/2021
Supplement Report FDA Received Date07/20/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ARC I2000SR INST, 03M74-02, ISR60755; ARC I2000SR INST, 03M74-02, ISR60755
-
-