| Model Number |
B2070-040 |
| Medical Device Problem Codes |
Off-Label Use (1494); Material Rupture (1546); Material Separation (1562)
|
| Health Effect - Clinical Codes |
Foreign Body In Patient (2687); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/17/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Device code (b)(4).The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was to treat a mildly tortuous, mildly calcified de novo subclavian vein that was 80% stenosed.A 7x40mm armada 35 balloon was inflated once and ruptured transversely at 15 atmospheres.Another same size armada 35 was used to successfully complete the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
B1, h1: type of reportable event changed from malfunction to serious injury
health effect impact code 2199 was removed.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was to treat a mildly tortuous, mildly calcified de novo subclavian vein that was 80% stenosed.A 7x40mm armada 35 balloon was inflated once and ruptured transversely at 15 atmospheres.Another same size armada 35 was used to successfully complete the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.Subsequent to the initially filed mdr report, device analysis noted the balloon and inner member were separated distal to the balloon proximal seal and the separated portions was not returned.The account confirmed that the separation occurred during the procedure and that the separated device portions were removed via snare.Angiography confirmed that nothing remained in the patient's anatomy.No additional information was provided.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was to treat a mildly tortuous, mildly calcified de novo subclavian vein that was 80% stenosed.A 7x40mm armada 35 balloon was inflated once and ruptured transversely at 15 atmospheres.Another same size armada 35 was used to successfully complete the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.Device analysis noted the balloon and inner member were separated distal to the balloon proximal seal and the separated portions were not returned.The account confirmed that the separation occurred during the procedure and that the separated device portions were removed via snare.Angiography confirmed that nothing remained in the patient¿s anatomy.Additional information was received by the account stating: the armada balloon was used to dilate the stenosed synthetic arteriovenous fistula graft instead of the central subclavian vein as initially reported.After the balloon ruptured at 15 atmospheres, the physician found no residual device in the patient's body via angiography.However, it was later confirmed that the armada 35 balloon had separated and the separated portions were never snared as previously reported and remain in the anatomy.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Visual and scanning electron microscopy (sem) inspection/analysis were performed on the returned device.The reported balloon rupture and separations were confirmed.A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The sem report determined the balloon exhibited a longitudinal rupture intersected by radial tearing.The origin area was not identified on the proximal half of the balloon separation and may possibly be along the longitudinal rupture on the distal half of the separation.The inner member exhibited a jagged edge, necking and bunching.The edge of the rupture exhibited tearing.The investigation determined the reported complaint and foreign body in patient appear to be related to operational context.There is no indication of a product quality issue with respect to design, manufacture or labeling of the device.Codes 4641, 4582, 1494 were removed.
|
| |
|