According to the reporter, during an open procedure, when suturing a biologic mesh, the 3 threads broke apart when pulling the knots.Another type of strand was used to resuture to resolve the issue.There was no patient injury.
|
|
Additional information: d9, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device and seventeen unused samples were returned for analysis.Visual inspection noted the opened sample had a suture and needle attached.Microscopic inspection of the suture end noted no breakage.For the sealed samples, a tactile and microscopic inspection of the sutures was performed with no abnormalities.The foils were inspected with light assisted microscopy with no abnormalities.Functional evaluation noted a tensile test was performed on the sealed samples and the results met the specifications for this product.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it has met all medtronic quality specifications.The reported condition was not confirmed.The investigation determined that the product tested within specifications and performed as intended.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|