• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. TWINFIX 2.8 TI W/NDL 22-0 ULTRABRAID; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. TWINFIX 2.8 TI W/NDL 22-0 ULTRABRAID; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE Back to Search Results
Model Number 72202067
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/22/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference (b)(4).
 
Event or Problem Description
It was reported that during a maxillofacial surgery the twinfix broke during use.The piece was retrieved using tweezers.The procedure was successfully completed without a significant delay using a back-up device.No other complications were reported.
 
Additional Manufacturer Narrative
The reported device, used in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection found a broken anchor tip with debris.An analysis of the customer provided image found the device with a broken anchor.Based on the condition of the product material found during visual inspection it was determined that additional material testing was not required.A review of the device records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.A review of risk management files found that the reported failure was documented appropriately.The instructions for use were reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A review of the material found that a statement (certification of compliance) or evidence of conformance to material and processing specifications must be provided.The complaint was confirmed.Factors that could have contributed to the reported event include excessive force or torque, off-axis insertion, or improper preparation of the insertion site.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TWINFIX 2.8 TI W/NDL 22-0 ULTRABRAID
Common Device Name
FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
MDR Report Key12160892
Report Number1219602-2021-01548
Device Sequence Number11827145
Product Code MBI
UDI-Device Identifier03596010619112
UDI-Public03596010619112
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K972326
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/25/2025
Device Model Number72202067
Device Catalogue Number72202067
Device Lot Number2058824
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/13/2021
Initial Date Received by Manufacturer 06/22/2021
Supplement Date Received by Manufacturer08/13/2021
Initial Report FDA Received Date07/13/2021
Supplement Report FDA Received Date08/16/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age56 YR
Patient Weight62
-
-