Model Number MC2-5070R
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code Failure of Implant (1924)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has been received and is pending evaluation.The root cause is unable to be determined at this time.A supplemental report will be submitted following the device evaluation.
 
Event or Problem Description
Information was received that the rod was not lengthening.No patient harm was reported.
 
Event or Problem Description
No additional information was received.
 
Additional Manufacturer Narrative
Device evaluation: the returned rod was visually inspected and no damage was found.Score marks observed on the distraction rod that confirmed the distraction rod was extended approximately 12 mm from its initial position.X-ray images of the internal components showed no damage and revealed the rod was partially distracted.The rod was functionally tested and was able to distract and retract with the erc and manual distractor.The length as received was measured at 228.65 mm.The minimum and maximum lengths were measured at 244.025 and 216.16 mm respectively; giving a total stroke measurement of 28.09mm, which meet the specification.Distraction force testing was measured at 45.50lbs, which meets the specification.The reported failure mode was unable to be confirmed as the rod was fully functional and met acceptance test specifications.Device records review: review of the device history record for the rod confirmed that it met all of the required quality inspections prior to release.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGEC SPINAL BRACING AND DISTRACTION SYSTEM
Common Device Name
GROWING ROD SYSTEM - MAGNETIC ACTUATION
Manufacturer (Section D)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise dr, suite 100
aliso viejo CA 92656
MDR Report Key12161111
Report Number3006179046-2021-00406
Device Sequence Number14502967
Product Code PGN
UDI-Device Identifier00887517834010
UDI-Public887517834010
Combination Product (Y/N)N
Initial Reporter StateAR
Initial Reporter CountryUS
PMA/510(K) Number
K171791
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMC2-5070R
Device Lot Number7090903
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/29/2021
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 06/23/2021
Supplement Date Received by Manufacturer08/26/2021
Initial Report FDA Received Date07/13/2021
Supplement Report FDA Received Date08/27/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age11 YR
Patient Weight37
-
-