| Model Number |
MC2-5070S |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has been received and is pending evaluation.The root cause is unable to be determined at this time.A supplemental report will be submitted following the evaluation.
|
| |
|
Event or Problem Description
|
|
Information was received that the rod was not lengthening.No patient harm was reported.
|
| |
|
Event or Problem Description
|
|
No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional data: h3, h6, h10 corrected data: d4 (lot number),g4, h6 (impact code & clinical code) device evaluation: visual inspection of the returned rod showed score marks on the distraction rod.The rod was unable to be distracted with the manual distractor and external remote controller (erc).Additionally, the rod was unable to be retracted with fast distractor.X-ray images of the internal components revealed a broken cylinder magnet.The reported failure was therefore confirmed as the rod was not able to be distracted or retracted.The exact root cause for the cylinder magnet fracture is unknown.Device records review: review of the device history records revealed no discrepancies related to this complaint.The rod was manufactured in accordance with the specified requirements and met all of the required quality inspections before shipment.
|
| |
|
Event or Problem Description
|
|
No additional information has been provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional device evaluation: measurements of the score marks confirmed the distraction rod was extended approximately 9mm from its initial position.The inspection data for the cylinder magnet (lot#171121-11) has been reviewed and confirmed the part met design specifications per the engineering drawings.Magnet fracture is a known inherent risk of the device.The exact root cause of the broken cylinder magnet is unable to be determined, however, the following possible root causes have been identified: the titanium casing was insufficient to protect magnet from loading; improper assembly of the magnet module resulting in unintended loading of the magnet; failure of the epoxy bond resulting in unintended loading of the magnet.
|
| |
|