| Catalog Number |
MX5301 |
| Medical Device Problem Code |
Difficult to Remove (1528)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/14/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Date of event: unknown.The date received by manufacturer has been used for this field.Investigation summary: one sample was received for quality investigation.The customer complaint of component damage could not be verified by investigation.The sample was first visually inspected for any physical defects or damage and nothing was observed.The connection between the male luer lock and the cannula that was submitted with the sample.The male luer connection was secure to the cannula and the connection was not loose.The extension set was then attached to a syringe and fluids were pushed through the tubing.There was no leakage at the luer connection, and there was no occlusion or air in line observed during the testing.A device history record review for model mx5301 lot number 21025131 was performed.The search showed that a total of 12,803 units in 1 lot number was built on (b)(6) 2021.There were no quality notifications issued for the failure mode reported by the customer during the production build of this set.Investigation conclusion: this incident has been added to our database of reported incidents.Our business team regularly reviews the collected data for identification of emerging trends.Root cause description: a root cause could not be determined due to the issue reported not being able to be replicated.
|
| |
|
Event or Problem Description
|
|
It was reported that 8-in pressure rated extension set was difficult to disconnect.The following information was provided by the initial reporter: it was reported by the customer that the spin on the male leur lock does not secure.Verbatim: the spin on the male leur lock does not seem to secure.When it does secure, the male leur is almost jammed in to the iv catheter hub and at times cannot be removed by hand.
|
| |
|
Search Alerts/Recalls
|