• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION VITALIO; IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION VITALIO; IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT) Back to Search Results
Model Number K273
Medical Device Problem Code Defective Device (2588)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 03/05/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A longevity calculation was completed and it was confirmed that the device did not meet longevity expectations.Device memory was reviewed and no faults or errors were noted.It was confirmed that the device was in ddd mode at the time the replacement indicator was declared.It was also noted that the device sensed in the atrial chamber 96% of the time while implanted with prolonged high atrial rates.The device was then exposed to simulated heart load conditions, and the pacing and sensing functions were tested.Impedance testing was completed and all measurements were within normal limits.The device operated appropriately with no interruptions in therapy output at the returned programmed settings.Chronic high atrial rates, as occurred with this device, can reduce longevity in devices that are programmed ddd(r) with atrial sensing on.Device power consumption increases proportional to the additional processing for sensed atrial events.As a result, the high rate of atrial sensing was determined to be the cause of the longevity being lower than expected.
 
Event or Problem Description
This product was returned to the manufacturer without any report of performance issues or adverse patient effects.The returned device was analyzed and the product investigation result indicated that this device exhibited chronic high atrial sensing which affected the longevity.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VITALIO
Common Device Name
IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
MDR Report Key12164000
Report Number2124215-2021-18917
Device Sequence Number4259841
Product Code LWP
UDI-Device Identifier00802526516962
UDI-Public00802526516962
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
N970003/S143
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 07/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date07/25/2015
Device Model NumberK273
Device Catalogue NumberK273
Device Lot Number100882
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/05/2021
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/12/2021
Initial Report FDA Received Date07/13/2021
Patient Sequence Number1
-
-