• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LEMAITRE VASCULAR, INC. OVER-THE-WIRE EMBOLECTOMY CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LEMAITRE VASCULAR, INC. OVER-THE-WIRE EMBOLECTOMY CATHETER Back to Search Results
Model Number 1651-44
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/14/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
We have not received the complaint device for evaluation.Hence, we could not conclusively determine the root cause of the malfunction.During the follow-up investigation, we were informed that the balloon ruptured while removing the clot.There was no balloon fragmentation as a result of this rupture that required any further intervention.The surgeon had inspected the device before use and he found it to be functional.The balloon was inflated with the recommended volume of saline.We were also informed that catheter was used in a stenotic vessel.Our review of the lot history records for this lot did not find any discrepancies either in the manufacturing or packaging process that could be related to this incident.Our ifu appropriately warns the users about the risks that could occur with the use of the embolectomy catheter including the risk of balloon rupture due to exposure to calcified plaque and overinflated balloon.We recommend exercising caution when encountering extremely diseased vessels.The arterial embolectomy catheter is not recommended for the removal of fibrous, adherent, or calcified material (e.G.Chronic clot, atherosclerotic plaque).This catheter is not designed to withstand the additional pull force needed to remove these materials.There was no harm to the patient as a result of this incident.The surgeon used a second over-the-wire embolectomy catheter to complete the procedure.
 
Event or Problem Description
The balloon of the lemaitre over-the-wire embolectomy catheter ruptured during an embolectomy procedure.There was no harm to the patient as the result of this incident.Surgeon used another over-the-wire embolectomy catheter to complete the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OVER-THE-WIRE EMBOLECTOMY CATHETER
Common Device Name
EMBOLECTOMY CATHETER
Manufacturer (Section D)
LEMAITRE VASCULAR, INC.
63 second ave
burlington MA 01803
MDR Report Key12165666
Report Number1220948-2021-00068
Device Sequence Number14602676
Product Code DXE
UDI-Device Identifier00840663100675
UDI-Public00840663100675
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K973334
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 07/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1651-44
Device Catalogue Number1651-44
Device Lot NumberOTW4738
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/14/2021
Initial Report FDA Received Date07/14/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-