• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: W. L. GORE & ASSOCIATES, INC. GORE® VIABAHN® ENDOPROSTHESIS WITH HEPARIN BIOACTIVE SURFACE; SYSTEM, ENDO GRAFT, ARTERIOVENOUS DIALYSIS ACCESS CIRCUIT STENOSIS TREATMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

W. L. GORE & ASSOCIATES, INC. GORE® VIABAHN® ENDOPROSTHESIS WITH HEPARIN BIOACTIVE SURFACE; SYSTEM, ENDO GRAFT, ARTERIOVENOUS DIALYSIS ACCESS CIRCUIT STENOSIS TREATMENT Back to Search Results
Model Number VBHR100502A
Medical Device Problem Code Activation Failure (3270)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/17/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Cbas® heparin surface incorporates cbas-heparin manufactured from heparin sodium api, which is covalently bound to the device surface and is essentially non-eluting.
 
Event or Problem Description
The following was reported to gore: on (b)(6) 2021, a patient presented with stenosis in the left basilic vein and underwent treatment utilizing a gore® viabahn® endoprosthesis with heparin bioactive surface.The physician was able to advance the device to the target lesion without issue using an 8fr sheath and 0.035" guidewire.When the physician attempted to deploy the device, the center of the device would not fully expand.The physician attempted to expand the device using post deployment ballooning, but was unsuccessful.Blood flow was confirmed through the device.The device remains implanted and the patient tolerated the procedure.
 
Additional Manufacturer Narrative
Added component code to section h6 corrected pma 510(k) number to p130006 in section g4.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® VIABAHN® ENDOPROSTHESIS WITH HEPARIN BIOACTIVE SURFACE
Common Device Name
SYSTEM, ENDO GRAFT, ARTERIOVENOUS DIALYSIS ACCESS CIRCUIT STENOSIS TREATMENT
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL ECHO RIDGE B/P
3250 w. kiltie lane
flagstaff AZ 86005
Manufacturer Contact
kaitlin barnash
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key12165985
Report Number2017233-2021-02163
Device Sequence Number17297118
Product Code PFV
UDI-Device Identifier00733132646029
UDI-Public00733132646029
Combination Product (Y/N)Y
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
P130006
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/09/2023
Device Model NumberVBHR100502A
Device Catalogue NumberVBHR100502A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/17/2021
Supplement Date Received by Manufacturer06/17/2021
Initial Report FDA Received Date07/14/2021
Supplement Report FDA Received Date03/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/09/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age61 YR
Patient SexMale
-
-