• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CS300; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CS300; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-3023-53
Device Problem Failure to Sense (1559)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/21/2021
Event Type  malfunction  
Manufacturer Narrative
Testing of actual/suspected device: (10/213) a getinge field service engineer (fse) evaluated the iabp but was unable to reproduce the reported "not sending signals to balloon" issue.The fse then performed all functional and safety checks to meet factory specifications.Unit passed all functional and safety test per factory specifications.The iabp was then released to the customer and cleared for clinical service.A supplemental report will be submitted upon completion of our investigation.The full event site name is (b)(6).
 
Event Description
It was reported that the cs300 intra-aortic balloon pump (iabp) was not sending signals to the balloon.It is unknown the circumstances under which the event occurred.It is also unknown if there was a patient involvement.However, there was no adverse event reported.
 
Event Description
N/a.
 
Manufacturer Narrative
Updated fields: b4,e1(site country),e2,e3, g3,g6,h2,h6(type of investigation, investigation findings, health effect ¿ impact codes, investigation conclusions), h10.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key12166217
MDR Text Key261490805
Report Number2249723-2021-01513
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567107882
UDI-Public10607567107882
Combination Product (y/n)N
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 02/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/14/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number0998-00-3023-53
Device Catalogue Number0998-00-3023-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Date Manufacturer Received02/20/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/01/2015
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
UNKNOWN.
Patient SexPrefer Not To Disclose
-
-