DEPUY IRELAND - 9616671 ATTUNE REVISION TIBIAL BLOCK R; RESECTION BLOCKS/DEVICES Back to Search Results
Model Number 2506-10-112
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/21/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the drill pin got stuck in revision attune tibial block right side.The disposal pin is stuck in the block and the packaging for the disposal pin was discarded.
 
Additional Manufacturer Narrative
Product complaint (b)(4).Investigation summary: the device associated with the reported event was not returned, however, review of the provided photo confirmed the reported event.Furthermore, the fixation pin appeared bent and twisted consistent with suspected off axis alignment of the mating pin under power resulting in the pin and block fusing together.Based on the observation of the investigation, the need for corrective action is not indicated.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: lot information not available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE REVISION TIBIAL BLOCK R
Common Device Name
RESECTION BLOCKS/DEVICES
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
MDR Report Key12166555
Report Number1818910-2021-15158
Device Sequence Number11712611
Product Code HTZ
UDI-Device Identifier10603295436232
UDI-Public10603295436232
Combination Product (Y/N)N
Initial Reporter CountryAS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2506-10-112
Device Catalogue Number250610112
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/23/2021
Supplement Date Received by Manufacturer09/13/2021
Initial Report FDA Received Date07/14/2021
Supplement Report FDA Received Date09/15/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age71 YR
Patient Weight95
-
-