• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AESCULAP INC TILTING BLADE CLAMP; GENERAL SURGICAL INSTRUMENTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AESCULAP INC TILTING BLADE CLAMP; GENERAL SURGICAL INSTRUMENTS Back to Search Results
Model Number MB510
Medical Device Problem Code Fitting Problem (2183)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/25/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Manufacturing site evaluation: additional information / investigation results will be provided in a supplemental report, if available.
 
Event or Problem Description
It was reported to aesculap inc.That atilting blade clamp (part # mb510) was used during an unknown procedure performed on (b)(6) 2021.According to the complainant,surgeon having issues with the ratchets fitting correctly on the rings.Recently, a ratchet came off and fell into a patient.Ratchet was retrieved with the help of blade.Additional information was not provided.The complaint device is available to be returned to the manufacturer for evaluation.The adverse event is filed under aic reference xc (b)(4).
 
Event or Problem Description
Update: the ring and tilting blade clamp were returned to the manufacturer.Associated medwatches: 2916714-2021-00139; 2916714-2021-00143.
 
Event or Problem Description
Associated medwatches: 2916714-2021-00143.
 
Additional Manufacturer Narrative
Investigation results: we received two additional table clamps and one round ring (mb522).The devices were sent back to the manufacturer for full investigation and a health hazard analysis.The manufacturer tested the ring with the clamps that were sent back and with current stock.The ring appeared to be slightly out of tolerance at one point on the ring but was conforming at all other points.It was determined that this minor deviation was not the cause of the reported error.The deviation caused the clamp to be stuck on the ring and difficult to move.Alternatively, if the deviation had caused the clamp to fit loosely on the ring, then a correlation could exist between the deviation and the clamp falling into the patient.Improper setup of the clamp onto the ring is more likely to have caused the device to fall off the ring and into the patient.The manufacturer determined that when these devices (mb510) are properly mounted to the ring (mb522), the reported error is not likely to be reproducible.Corrective and preventive actions: 1.Upon receiving notification of the reported complaint, mb510 and mb522 were held for containment and not sold until the closure of this investigation.2.A trending analysis was conducted and indicated the following: since 2009, there were over 20,000 clamps (mb510) and over 300 rings (mb522) sold, and none were returned with the reported error.3.A health hazard analysis was conducted, and the results indicated that the reported error was not likely to occur and that the potential risk if it did occur to be limited/ transient.Since the root cause analysis indicated that the cause of the reported error was not manufacturing-related, the need for a mdr or product recall was ruled out.4.To address the out of tolerance issue, the manufacturer updated their inspection protocols.Their current protocols only required one point to be tested during final inspection.Their updated protocol calls for four different test points during final inspection.Device disposition: the manufacturer reworked the ring to fix the deviation that was found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TILTING BLADE CLAMP
Common Device Name
GENERAL SURGICAL INSTRUMENTS
Manufacturer (Section D)
AESCULAP INC
3773 corporate parkway
center valley PA 18034
Manufacturer (Section G)
AESCULAP INC
3773 corporate parkway
center valley PA 18034
Manufacturer Contact
lindsay chromiak
3773 corporate parkway
center valley, PA 18034
MDR Report Key12166897
Report Number2916714-2021-00139
Device Sequence Number14606297
Product Code GAD
UDI-Device Identifier04046964567568
UDI-Public4046964567568
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMB510
Device Catalogue NumberMB510
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/10/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/25/2021
Supplement Date Received by Manufacturer07/08/2021
11/16/2021
Initial Report FDA Received Date07/14/2021
Supplement Report FDA Received Date08/02/2021
12/13/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-