|
SMITH & NEPHEW, INC. DISTAL CUT BLOCK 5.5MM; PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER
|
Back to Search Results |
|
| Model Number |
74013785 |
| Medical Device Problem Code |
Corroded (1131)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/06/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that distal cut block (4.5mm to 8.5mm) are rusted.As this was noticed upon field inspection, there was not patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
H3, h6: the device, intended for use in treatment, was returned for evaluation.A visual inspection of the returned device confirmed the stated failure mode.The device was found to have discoloration on it, rendering the device in operable.The device shows signs of extensive use.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal any manufacturing or material abnormalities that could have caused or contributed to the reported incident.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|
Search Alerts/Recalls
|
|
|