Model Number 71631140
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/06/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a internal fixation surgery, a guide bolt wrench was broken.This happened while removing the guide bolt from the nail with the wrench after inserting a nail.The wrench did not rotate, surgeon hit it with a hammer and the wrench broke off at the constricted part.The tip of the device remained inside the guide bolt was removed.The inserted nail and screw was removed, then the guide bolt and nail were separated.The removed nail was inserted again.Surgery was resume, with smith and nephew surgical delay, with a smith and nephew back-up device and unknown competitor device.It is unknown if the patient was not harmed as consequence of this problem.
 
Additional Manufacturer Narrative
H3, h6: the associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The device was broken along the tip, rendering the device inoperative.The device shows signs of extensive use.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GUIDE BOLT WRENCH
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key12168090
Report Number1020279-2021-05966
Device Sequence Number17197476
Product Code HSB
UDI-Device Identifier03596010419644
UDI-Public03596010419644
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K122170
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71631140
Device Catalogue Number71631140
Device Lot Number03HM02959
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/09/2021
Initial Date Received by Manufacturer 07/06/2021
Supplement Date Received by Manufacturer08/18/2021
Initial Report FDA Received Date07/14/2021
Supplement Report FDA Received Date08/19/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-