| Model Number |
ONB12SHF |
| Medical Device Problem Codes |
Detachment of Device or Device Component (2907); Packaging Problem (3007)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/11/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was not returned, but a photo was available for evaluation.Visual inspection noted sharp damage on the trocar circular seal.It was reported that there was a damage to a seal in the device and the seal disengaged from the port.The reported issues were confirmed.The product analysis noted evidence that the device was not used as intended.This issue can occur when contact is made with a sharp surgical instrument during clinical application.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: use special care when introducing or removing sharp-edged or sharp-angled endoscopic instruments to minimize the potential of inadvertent damage to the seal if information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a laparoscopic nephrectomy, the seal part got damaged and about 2mmx2mm of broken piece dropped into the abdominal cavity.The device was used as it was.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted sharp damage on the trocar circular seal.Functionally, the device passed an air leak test.It was reported that there was damage to a seal in the device, and the seal disengaged from the port.The reported issues were confirmed.The product analysis noted evidence that the device was not used as intended.Cut in the circular seal may occur when contact is made with a sharp surgical instrument during clinical application.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: use special care when introducing or removing sharp-edged or sharp-angled endoscopic instruments to minimize the potential of inadve rtent damage to the seal.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|