| Catalog Number |
320617 |
| Medical Device Problem Code |
Failure to Deliver (2338)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/15/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation summary: no samples (including photos) were returned therefore the complaint could not be confirmed and the root cause is undetermined.This is the 2nd complaint for the reported lot number.A review of the manufacturing records was performed and no non-conformances were raised in association with this type of event for this lot.Complaints received for this device and reported condition will continue to be tracked and trended.If samples are received in the future the complaint will be reopened for further investigation.Investigation conclusion: lot#0140339 dhr was reviewed and no qn found.Manufacture records were reviewed, no abnormality was found.Pen needle product has 100% clog test in flowguru station, and defect part will be rejected by flowguru station automatically.Clog test was conducted on 7pcs retention samples and all no needle clog fail found.No samples were returned for further investigation.Base on investigation above, the certain cause cannot be concluded at the moment.
|
| |
|
Event or Problem Description
|
|
It was reported that the pn relion 31g x 6mm 3b was unable to deliver insulin.The following information was provided by the initial reporter: consumer reported finding 1 pen needle to clog during injection.Does not complete flow check before injections.Changed this pen needle from baslagar pen and completed injections with another pen needles.
|
| |
|