• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. RUBY COIL LP; HCG, KRD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. RUBY COIL LP; HCG, KRD Back to Search Results
Model Number RBYLP0860
Medical Device Problem Code Failure to Advance (2524)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/15/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report numbers: 3005168196-2021-01543, 3005168196-2021-01544.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure to treat a carotid cavernous fistula (ccf) using penumbra smart coils (smart coil), a ruby coil lp, a packing coil lp, a benchmark bmx96 access system (benchmark max) and a non-penumbra microcatheter.During the procedure, the physician placed a benchmark max in the right jugular vein.The physician then placed two smart coils into the carotid cavernous sinus using the microcatheter.Afterwards, a ruby coil lp and a smart coil would not advance at all within their respective introducer sheaths; therefore, the ruby coil lp and the smart coil were removed.Next, the physician experienced resistance while advancing approximately ten centimeters of a packing coil lp through the distal end of the microcatheter.Subsequently, while attempting to retract the packing coil lp, the physician experienced resistance and heard a popping sound.The physician then removed the pusher assembly of the packing coil lp and noticed the coil broke and unraveled.Afterwards, the microcatheter was removed and it was noticed that the coil was hanging from the hub of the microcatheter.Subsequently, while the physician attempted to pull on the coil, it was noticed that the previously placed coil mass was also moving, and the packing coil lp would not move in the carotid cavernous sinus.Consequently, the physician flushed the coil into the jugular vein.Angiographic runs indicated that the fistula embolized with the two smart coils that were previously placed.The coil did not appear to be restricting flow in the jugular vein.The procedure ended at this point.There was no report of an adverse effect to the patient.
 
Additional Manufacturer Narrative
Please note that the following section was incorrectly reported on the initial mfr report and is being corrected on this follow-up #01 mfr report 3005168196-2021-01542.1.Section b.Box 5.Describe event or problem this report is associated with mfr report numbers: 1.3005168196-2021-01543.2.3005168196-2021-01544.H3 other text : placeholder.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure to treat a carotid cavernous fistula (ccf) using penumbra smart coils (smart coil), a ruby coil lp, a packing coil lp, a benchmark bmx96 access system (benchmark max) and a non-penumbra microcatheter.During the procedure, the physician placed a benchmark max in the right jugular vein.The physician then placed two smart coils into the carotid cavernous sinus using the microcatheter.Afterwards, a ruby coil lp and a smart coil would not advance at all within their respective introducer sheaths; therefore, the ruby coil lp and the smart coil were removed.Next, the physician experienced resistance while advancing approximately ten centimeters of a packing coil lp through the distal end of the microcatheter.Subsequently, while attempting to retract the packing coil lp, the physician experienced resistance and heard a popping sound.The physician then removed the pusher assembly of the packing coil lp and noticed the coil broke and unraveled.Afterwards, the microcatheter was removed and it was noticed that the remaining coil was hanging from the hub of the benchmark max.Subsequently, while the physician attempted to pull on the coil, it was noticed that the previously placed coil mass was also moving, and the packing coil lp would not move in the carotid cavernous sinus.Consequently, the physician flushed the coil into the jugular vein.Angiographic runs indicated that the fistula embolized with the two smart coils that were previously placed.The coil did not appear to be restricting flow in the jugular vein.The procedure ended at this point.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RUBY COIL LP
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key12171168
Report Number3005168196-2021-01542
Device Sequence Number16524391
Product Code HCG
UDI-Device Identifier00815948021884
UDI-Public00815948021884
Combination Product (Y/N)Y
Initial Reporter CountryCA
PMA/510(K) Number
K192955
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/13/2021
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/21/2026
Device Model NumberRBYLP0860
Device Catalogue NumberRBYLP0860
Device Lot NumberF104029
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 06/15/2021
Supplement Date Received by Manufacturer01/14/2005
Initial Report FDA Received Date07/15/2021
Supplement Report FDA Received Date10/14/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age71 YR
Date Report Sent to FDA01/01/2005
-
-