| Model Number |
CNX500T17 |
| Medical Device Problem Code |
Degraded (1153)
|
| Health Effect - Clinical Code |
Nodule (4551)
|
| Date of Event |
07/05/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The manufacturer received information alleging a continuous positive airway pressure (cpap) device's sound abatement foam became degraded and caused the patient to have lung nodules.There is no report of the medical intervention that the patient has received at this time the manufacturer's investigation is ongoing.A follow-up report will be submitted when the manufacturer's investigation is complete.
|
| |
|
Additional Manufacturer Narrative
|
|
The manufacturer previously reported an issue related to sound abatement foam.This action was reported to fda per 21 cfr part 806.The device will be corrected per res 88058.
|
| |
|
Event or Problem Description
|
|
The manufacturer received information alleging a continuous positive airway pressure (cpap) device's sound abatement foam became degraded and caused the patient to have lung nodules.There is no report of the medical intervention that the patient has received at this time.This issue was reported to the fda per 21 cfr 806.The device will be corrected per res 88058.
|
| |
|
Additional Manufacturer Narrative
|
|
The manufacturer previously received information alleging a continuous positive airway pressure (cpap) device's sound abatement foam became degraded and caused the patient to have lung nodules.There is no report of the medical intervention that the patient has received at this time.There was no report of serious patient harm or injury.Section b1 was corrected for product problem.(both adverse event and product problem was checked in the previous mdr.) section b2 was corrected to blank.(previously, it was other serious or important medical events.) section h1 was changed from serious injury to malfunction.Section h6 health effects - impact code was corrected, type of investigation findings and investigation conclusions has been updated.The device was returned to manufacturer's service center for further evaluation.During the evaluation of the device at the manufacturer's service centre, the device was visually inspected and found no evidence of foam degradation.The manufacturer's service center concludes that they couldn't confirm the customer's allegation and there was no visibl foam degradation and unit passed the final test.
|
| |
|