| Model Number |
VCP496H |
| Medical Device Problem Code |
Material Frayed (1262)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/21/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Component code: (b)(4).This is a combination product, and the event has been reviewed for both the suture and the triclosan.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Additional information was requested and the following was obtained: please provide lot number: rd2akl.Trade name: irgacare¿, active ingredient(s): triclosan, dosage form: suture/solid/parenteral, strength: 275 g/m.Events reported in 2210968-2021-06364.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent a davinci adrenalectomy on (b)(6) 2021 and suture was used.During the procedure, the sutures debraided and disintegrated while tying.Another like device was used to complete the procedure.No adverse patient consequences were reported.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Date sent to the fda: 8/05/2021.H6 component code: g07002 - no device problem found.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc., or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H3: evaluation: h3 investigation summary: one packet of product code vcp496h were returned for analysis with the packaging closed.Upon initial inspection to the sample no external damages were observed on the packet.In order to evaluate the conditions of the returned sample, the packet was opened, and no defects were detected.The swage and attachment area were noted to be as expected.The suture was dispensed without problems and examined along of the strand to detect any issue related to fraying or damaged suture and no defects were observed during evaluation.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.It should be noted that as part of our quality process, each batch is randomly inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.There may have been other circumstances or issues that occurred during the use of the device that we were unable to duplicate during our laboratory analysis.A manufacturing record evaluation was performed for the finished device, and no non-conformances were identified.
|
| |
|
Search Alerts/Recalls
|
|
|