MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 HUBLESS FLAT SILICONE DRAIN, XRAY OPAQUE STRIPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 HUBLESS FLAT SILICONE DRAIN, XRAY OPAQUE STRIPE Back to Search Results
Model Number 0070430
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Codes Foreign Body In Patient (2687); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 04/22/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
 
Event or Problem Description
It was reported that patient underwent a fusion and fixation of l2 & l3 surgical procedure on (b)(6) 2021.Upon post-operative follow-up on (b)(6) 2021 revealed a retained foreign object (part of the drain pulled on the day of discharge) via imaging.The remaining was surgically removed on (b)(6) 2021.The patient retained a significant portion of the drain on (b)(6).
 
Additional Manufacturer Narrative
The reported event was confirmed however the cause was unknown.One sample was confirmed to exhibit the reported failure.The product was used for the treatment.The product had caused the reported failure.Visual evaluation of the returned sample noted one opened (without original packaging) used flat silicone drain.Visual inspection noted a piece of the drain was broken off.A potential root cause for this failure mode could be due to incorrect cleaning method.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "important" a.Check for fluid entering reliavac® evacuator.Lack of flow may indicate: - all exudate has been removed.- wound drain is clogged and may require irrigation and aspiration (consult physician).- auxiliary wall suction pressure is above 210mm hg.- deflated balloon: check all connections for air leak and wound tube perforations for exposure above the skin.If still deflated, replace evacuator.B.When not using auxiliary suction during surgical wound closure, several activations of the reliavac® evacuator may be required to establish suction because of: - air entering partially closed wound.- an operative air pocket.C.Insert safety pin into hole in collar to attach evacuator to patient's clothing or bed linen.Caution: federal (u.S.A.) law restricts this device to sale by or on the order of a physician.To avoid the possibility of a hematoma due to wound evacuation, the instructions for use should be carefully followed.To avoid the possibility of drain damage or breakage: additional perforations should not be made in the drains.Avoid suturing through drains.Drains should lie flat and in line with the skin exit areas.Particular care should be taken to avoid any obstacles to the drain exit path.Drains should be checked during closure for free motion to avoid possibility of breakage.Drain removal should be done gently by hand.They should not be handled with pointed, toothed or sharp instruments which could cause cuts or nicks and lead to subsequent structural failure of the drain.Warning: surgical removal may be necessary if drain is difficult to remove or breaks.Indications for use: wound drains are used to remove exudates from wound sites contraindications: do not use for chest drainage.Warning: do not bypass/inactivate anti-reflux valve.Not made with natural rubber latex.Sterilized by ethylene oxide single use.Sterile unless package is opened or damaged.Do not use if package is opened or damaged.Do not resterilize.Caution: federal (u.S.A.) law restricts this device to sale by or on the order of a physician.A.Use with single flat drain - 1.Place perforated wound drain within critical fluid collection area of wound.2.Draw drain tube through skin/stab wound incision until flat portion of drain is seated appropriately.3.Trim drain tube to desired length and attach to blue adapter.Connect other end of blue adapter to y-connector.4.Insert connecting tube in reliavac® port a, up to indicator ring.B.Use with two flat drains - 1.Place perforated portion of wound drains within critical fluid collection areas of wound.2.Draw drain tubes through skin/stab wound incision until flat portion of the drains are seated appropriately.3.Trim drain tubes to desired length.4.Cut off plug from closed arm of y-connector and attach blue adapters.5.Attach drains to blue adapters.6.Insert connecting tube in reliavac® port a, up to indicator ring.Caution: punctures or additional perforations should not be made in the silicone wound drain.C.Attaching to auxiliary suction : 1.Insert suction adapter into port b.2.During auxiliary suction, balloon will inflate and exudate will flow over balloon surface from port a to port b.3.To discontinue auxiliary suction, remove suction adapter and close port b.Caution: do not use with wall suction in excess of 210mm hg.D.To establish suction : 1.Open port b.2.Pump bulb until balloon fills container.3.Close port b.Note: hissing sound is normal and stops when maximum suction pressure is reached.Possible reflux of fluid to the patient is reduced during reliavac® evacuator reactivation by a built-in anti-reflux valve in port a.E.To empty container : 1.Open port b.2.Invert unit.3.Pump bulb to empty quickly.F.To re-establish suction - 1.Repeat step "d" above.G.To read fluid volume - 1.Open port b.2.Allow balloon to deflate.3.Read and record volume.4.To reactivate, repeat step "d" above." h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that the fusion and fixation of l2 and l3 surgical procedure on (b)(6) 2021.Upon post operative follow-up on (b)(6) 2021 it was revealed via imaging that a part of the drain pulled on the day of discharge was retained.The remaining part was surgically removed on (b)(6) 2021.It was stated that the patient retained a significant portion of the drain on (b)(6) 2021.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUBLESS FLAT SILICONE DRAIN, XRAY OPAQUE STRIPE
Common Device Name
SILICONE DRAIN
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
yonic anderson
8195 industrial blvd
covington 30014
7707846100
MDR Report Key12174903
Report Number1018233-2021-04249
Device Sequence Number13836209
Product Code GBX
UDI-Device Identifier00801741090752
UDI-Public(01)00801741090752
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date10/31/2025
Device Model Number0070430
Device Catalogue Number0070430
Device Lot NumberNGEX1423
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/22/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/22/2021
Supplement Date Received by Manufacturer12/29/2021
Initial Report FDA Received Date07/15/2021
Supplement Report FDA Received Date01/16/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/03/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-