| Model Number |
QD8-S |
| Medical Device Problem Codes |
Vibration (1674); Device-Device Incompatibility (2919)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.The actual device was returned for evaluation.The attachment device was evaluated and the reported condition that the device was damaged and would not hold the drill in place while drilling was not confirmed.Therefore, an assignable root cause was not determined.However, during evaluation it was observed that the device had corrosion/rusting/pitting, and vibration.It was further determined that the device failed pretest for vibration.The assignable root cause of this condition was determined to be traced to maintenance, which is user error/misuse/abuse.Udi: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the attachment device was damaged and would not hold the drill in place while drilling.During in-house engineering evaluation it was determined that the device had vibration.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention, or prolonged hospitalization.The exact date of the event was not reported.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
| |
|