Model Number AAB00
Medical Device Problem Codes Break (1069); Scratched Material (3020)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/30/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Corrected data: this event was initially included in vmsr filing 2648035-2021-00010.Due to the inclusion of other ancillary details that may appear as causes in the vmsr filing, this event is now being captured as an initial 30 day mdr.This report captures the event for serial number (b)(4).This is report number 1 out of 2 reports that are being submitted as initial individual mdrs.Date of birth or age at time of event: unknown/not provided.Sex: unknown/not provided.If explanted, give date: not applicable iol was not explanted.Phone: (b)(6).Email address: unknown, information not provided.Device evaluation: product testing could not be performed as the product was not returned.The reported event could not be verified, and no product deficiency could be identified.Manufacturing records review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specifications.A search revealed that no additional complaints were received for this production order.Conclusion: based on the investigation, no product deficiency was identified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that the surgeon checked the iol before the surgery.However, after implanting the iol into the patient¿s operative eye, he found that some "curve" on the surface of the iol.Patient didn¿t have any problematic reaction and the iol remains implanted.The surgeon is not planning any surgical intervention.No further information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSAR
Common Device Name
MONOFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
1700 e st andrew place
santa ana CA 92705
MDR Report Key12175865
Report Number2648035-2021-00079
Device Sequence Number1521854
Product Code HQL
UDI-Device Identifier05050474548725
UDI-Public(01)05050474548725(17)231122
Combination Product (Y/N)Y
Initial Reporter CountryTW
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/22/2023
Device Model NumberAAB00
Device Catalogue NumberAAB0000230
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/15/2021
Initial Report FDA Received Date07/15/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-