COOK INC CXI SUPPORT CATHETER; KRA CATHETER, CONTINUOUS FLUSH Back to Search Results
Model Number G56553
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/28/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.
 
Event or Problem Description
As reported, during an unknown procedure, the tip of a cxi support catheter separated.Additional information has been requested.
 
Event or Problem Description
Additional information was received 27aug2021.The device did not make patient contact.The tip separated as the user was loading the catheter over an unknown wire guide.The patient's condition was described as stable.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
There is no new patient or event information to report.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unchanged, or unavailable.Description of event: as reported, during an unknown procedure, the tip of a cxi support catheter separated.The device did not make patient contact.The tip separated as the user was loading the catheter over an unknown wire guide.The patient's condition was described as stable.Investigation ¿ evaluation: reviews of the complaint history, device history record, drawing, instructions for use (ifu), manufacturing instructions, and quality control procedures were conducted during the investigation.A visual inspection of the complaint device and unused product was also conducted.The complainant returned one prior to use device and nine sealed, unused devices to cook for investigation.The catheter that was opened had a tip length that measured below specification.On the sealed devices, the tip length measured within specification.The sealed devices were loaded on a sample 0.018" wire via the distal tip and again via proximal hub.No tip breakage was noted.A document-based investigation evaluation was performed.One relevant subassembly nonconformance was noted; however, the affected device was scrapped.There have been two additional related complaints on the lot for the same failure mode, reported by the same user facility.The instructions for use (ifu) instructs, ¿upon removal from package, inspect the product to ensure no damage has occurred.¿ the information provided upon review of the dmr, dhr, ifu, and investigation of return device suggests that it is possible that the device was manufactured out of specification and that there are non-conforming devices in-house or in the field.However, due to the nature of the manufacturing process, it is unlikely that the tip deficiency affected the entire lot.Based on the information provided and the results of the investigation, cook has concluded that a manufacturing deficiency at the subassembly level likely contributed to the failure mode.The complaint lot was placed on stop ship and scrapped.Re-training of appropriate personnel has been completed.The risk analysis for the failure mode was reviewed, and it was determined that no escalation actions were required.The appropriate personnel have been notified and cook will continue to monitor for similar events.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CXI SUPPORT CATHETER
Common Device Name
KRA CATHETER, CONTINUOUS FLUSH
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key12176135
Report Number1820334-2021-01819
Device Sequence Number887324
Product Code KRA
UDI-Device Identifier00827002565531
UDI-Public(01)00827002565531(17)240415(10)13895156
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K122796
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/15/2024
Device Model NumberG56553
Device Catalogue NumberCXI-2.6-18-150-ANG
Device Lot Number13895156
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 06/30/2021
Supplement Date Received by Manufacturer08/27/2021
12/14/2021
Initial Report FDA Received Date07/15/2021
Supplement Report FDA Received Date09/15/2021
01/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/15/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-