| Model Number |
86200 |
| Medical Device Problem Codes |
Unsealed Device Packaging (1444); Material Puncture/Hole (1504)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/23/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a ultra ice plus was selected to be used in a procedure.However upon opening the box the sterile packageing was in an opened state.The device was replaced and the procedure was completed.
|
| |
|
Additional Manufacturer Narrative
|
|
Visual inspection of the device showed no damage related to the event was encountered during the visual inspection of the device.The packaging was not returned there for analysis of the sterile bag could not be performed.However media inspection was performed to the pictures sent to boston scientific, in which showed a tear of the sterile bag, thus, confirming the reported allegation.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.
|
| |
|
Event or Problem Description
|
|
It was reported that a ultra ice plus was selected to be used in a procedure.However upon opening the box the sterile packaging was in an opened state.The device was replaced and the procedure was completed.
|
| |
|