Visual inspection of the device showed no damage related to the event was encountered during the visual inspection of the device.The packaging was not returned there for analysis of the sterile bag could not be performed.However media inspection was performed to the pictures sent to boston scientific, in which showed a tear of the sterile bag, thus, confirming the reported allegation.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.
|