Model Number AAB00
Medical Device Problem Codes Break (1069); Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Corrected data: this event was initially included in vmsr filing 2648035-2021-00010.Due to the inclusion of other ancillary details that may appear as causes in the vmsr filing, this event is now being captured as an initial 30 day mdr.This report captures the event for serial number (b)(4).This is report number 2 out of 2 reports that are being submitted as initial individual mdrs.Date of birth or age at time of event: unknown/not provided.Sex: unknown/not provided.Date of event: unknown/not provided.If implanted, give date: not applicable iol was not implanted.If explanted, give date: not applicable iol was not explanted.Device evaluation: product testing could not be performed as the product was not returned.The reported event could not be verified, and no product deficiency could be identified.Manufacturing records review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specifications.A search revealed that no additional complaints were received for this production order.Conclusion: based on the investigation, no product deficiency was identified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that the iol ¿crumpled¿ during the second advancement.The surgeon deemed the inserter as defective.The procedure was successfully completed with another iol of the same model.There was no surgical intervention required.Reportedly, the patient had problems after being discharged however the problems were not specified.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSAR
Common Device Name
MONOFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
1700 e st andrew place
santa ana CA 92705
MDR Report Key12176853
Report Number2648035-2021-00080
Device Sequence Number11143421
Product Code HQL
UDI-Device Identifier05050474548718
UDI-Public(01)05050474548718(17)230618
Combination Product (Y/N)Y
Initial Reporter CountryCA
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 07/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/18/2023
Device Model NumberAAB00
Device Catalogue NumberAAB0000225
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/15/2021
Initial Report FDA Received Date07/15/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-