| Model Number |
03.404.017S |
| Medical Device Problem Codes |
Break (1069); Entrapment of Device (1212)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
06/16/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Additional procode: hrx.Complainant part is expected to be returned for manufacturer review/ investigation, but has yet to be received.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2021 during a femoral procedure due to osteomyelitis infarction, one (1) 10.5mm reamer head and one (1) 11.0mm reamer head tines were broken and embedded in the patient.One (1) drive shaft was broken during reaming.There was a surgical delay of fifteen (15) minutes.The procedure was successfully completed.There was patient consequence.Concomitant device reported: unknown reaming rod (part # unknown; lot # unknown; quantity: 1), unknown drill bits (part # unknown; lot # unknown; quantity: unknown).This report is for one (1) 10.5mm reamer head for ria 2 sterile.This is report 1 of 3 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: b6; h7; h9.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.D7a: unknown h5.
|
| |
|
Search Alerts/Recalls
|