• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DAVOL INC., SUB. C.R. BARD, INC. -1213643 MODIFIED ONFLEX; SURGICAL MESH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVOL INC., SUB. C.R. BARD, INC. -1213643 MODIFIED ONFLEX; SURGICAL MESH Back to Search Results
Catalog Number 0115610
Medical Device Problem Code Patient-Device Incompatibility (2682)
Health Effect - Clinical Codes Hypersensitivity/Allergic reaction (1907); Urticaria (2278)
Date of Event 07/02/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
As reported, the patient developed an allergic reaction and urticaria post-implant of the bard/davol modified onflex mesh.No product identifiers have been provided; therefore, a review of the manufacturing records is not possible.Based on the information available, no conclusions can be made.Should additional information be provided, a supplemental mdr will be submitted.
 
Event or Problem Description
As reported, following implant of a bard/davol modified onflex mesh, the patient experienced an allergic reaction and developed urticaria on (b)(6) 2021.As alleged, the patient was scheduled for a hospital visit on (b)(6) 2021 and the mesh is allegedly scheduled to be removed on (b)(6) 2021.
 
Event or Problem Description
As reported, following implant of a bard/davol modified onflex mesh, the patient experienced an allergic reaction and developed urticaria on (b)(6) 2021.As alleged, the patient was scheduled for a hospital visit on (b)(6) 2021 and the mesh is allegedly scheduled to be removed on (b)(6) 2021.Addendum: as reported, the patient underwent reactivity testing.As reported, the patient was allergic to iodine, negative to the mesh and slightly allergic to the absorbable ring.It was reported that the mesh removal was scheduled on (b)(6) 2021; the surgery has been postponed (time undecided) because the patient¿s allergic reaction symptoms (urticaria, etc.) improved before the date.The patient is under observation.
 
Additional Manufacturer Narrative
As reported, the patient developed an allergic reaction and urticaria post-implant of the bard/davol modified onflex mesh.No product identifiers have been provided; therefore, a review of the manufacturing records is not possible.Based on the information available, no conclusions can be made.Addendum to the initial mdr submitted.Additional information provided states that the patient underwent reactivity testing, and was found to be allergic to iodine, and slightly allergic to the absorbable ring component within the mesh implant.As reported, the patient had no reaction to the mesh component of the implant.As reported, the patient has been under observation and their symptoms have been improving, therefore the initially scheduled explant procedure has not been undertaken.Based on the additional information received, the cause of the patient¿s complication appears to be related to a reaction to iodine and the implant¿s absorbable ring.Regarding the patient¿s reaction to the implant¿s absorbable ring, the ifu supplied with the device states that the absorbable ring ¿fully degrades in vivo by means of hydrolysis and has been found to elicit an inflammatory response during absorption.¿ should additional information be provided a supplemental mdr will be submitted.Note: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : not returned - remains implanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MODIFIED ONFLEX
Common Device Name
SURGICAL MESH
Manufacturer (Section D)
DAVOL INC., SUB. C.R. BARD, INC. -1213643
100 crossings blvd.
warwick RI 02886
MDR Report Key12177526
Report Number1213643-2021-20185
Device Sequence Number1521865
Product Code FTL
UDI-Device Identifier00801741095399
UDI-Public(01)00801741095399
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K142706
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number0115610
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/02/2021
Supplement Date Received by Manufacturer07/27/2021
Initial Report FDA Received Date07/15/2021
Supplement Report FDA Received Date08/17/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
-
-