Model Number MMT-105ELPKNA
Device Problem Patient Device Interaction Problem (4001)
Patient Problem Hyperglycemia (1905)
Event Date 07/12/2021
Event Type  malfunction  
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event Description
The customer's family member reported via phone call that no insulin when the button is pushed.Customer's family member primed the insulin pen multiple times and replaced the needles, but the issue was not resolved.Customer stated the screw pulls back into the pen while dialing.Customer¿s blood glucose was 200 mg/dl.No harm requiring medical intervention was reported.The insulin pen will be returned for analysis.
 
Manufacturer Narrative
The screw was not bent, advanced when dosage knob was pressed dialing a dosage and retracted appropriately.No resistance was observed when dosing without a cartridge installed.The screw advanced every time 30.0 units was dialed and dosed until the screw reached max extension.No lead screw anomalies noted.No physical damage to injection foot or inpen was noted.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INPEN MMT-105ELPKNA ELI LILY PINK
Type of Device
SYRINGE, PISTON
Manufacturer (Section D)
COMPANION MEDICAL INC
1223 world trade drive, 100
san diego CA 92128
MDR Report Key12178497
MDR Text Key263848307
Report Number3012822846-2021-00592
Device Sequence Number1
Product Code FMF
UDI-Device Identifier000010862088000337
UDI-Public(01)000010862088000337(17)220930
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,other
Type of Report Initial,Followup
Report Date 08/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date09/30/2022
Device Model NumberMMT-105ELPKNA
Device Catalogue NumberMMT-105ELPKNA
Device Lot NumberB93VA
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/02/2021
Initial Date Manufacturer Received 07/12/2021
Initial Date FDA Received07/15/2021
Supplement Dates Manufacturer Received08/10/2021
Supplement Dates FDA Received08/11/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age17 YR
Patient Weight160
-
-