CORCYM S.R.L. CARBOMEDICS STANDARD PROSTHETIC HEART VALVE, MITRAL; MECHANICAL HEART VALVE PROSTHESIS
|
Back to Search Results |
|
| Model Number |
CPHV |
| Medical Device Problem Code |
Difficult to Open or Close (2921)
|
| Health Effect - Clinical Codes |
Pneumonia (2011); Septic Shock (2068); Unspecified Kidney or Urinary Problem (4503)
|
| Date of Event |
06/15/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
On (b)(6) 2021, a patient received a carbomedics standard mitral valve.Upon bypass weaning, a mitral valve prosthesis dysfunction was noted as also confirmed on arrival at icu with transesophageal echocardiogram.The patient presented acute prosthesis dysfunction, with a fixed disc, for which a mitral valve replacement was again performed, currently complicated with pneumonic process, acute kidney injury and resolution of septic shock over time with mechanical ventilation.The device was replaced with a biological valve perimount mitral 31.
|
| |
|
Additional Manufacturer Narrative
|
|
Fields updated: b4, g3, g6, h1, h2, h6.A complete manufacturing and material records review for the device has been performed.The results confirmed that the device satisfied all material, visual and performance standards required at the time of manufacture and release.Further tests will be performed upon device receipt.The conclusions are not yet available.
|
| |
|
Additional Manufacturer Narrative
|
|
Fields updated: b4, d9, g3, g6, h1, h2, h6.The device was not returned despite follow-ups; thus, no further investigation on the explanted device could be performed.Based on the available information, the root cause of the event could not be established.Based on the information received, the standard mitral valve size 33 was explanted and perimount valve size 31 was implanted instead.As such, the cause of the reported event could possibly be related to missizing but this cannot ultimately be confirmed since the device was not returned to the manufacturer for further investigation.However, no deficits were noted during the production records review.Should further information and/or the device be received in the future, the manufacturer will re-visit the case and take further actions as applicable.
|
| |
|