• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTECH SA ROSA ONE; COMPUTER-ASSISTED SURGICAL DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTECH SA ROSA ONE; COMPUTER-ASSISTED SURGICAL DEVICE Back to Search Results
Model Number ROSA ONE
Device Problem Patient Data Problem (3197)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/23/2021
Event Type  malfunction  
Manufacturer Narrative
A full analysis of the data logs and the patient folder has been performed and this analysis did not allow for any conclusion on the impossibility to open the patient folder on the device.The error could not be reproduced except by manually editing the.Hdr files.It is still possible that some operation on the patient folder modified the.Hdr files but in this case the files should not have been loaded on the maintenance session.Unique identifier (udi) #: (b)(4).
 
Event Description
It was reported that upon attempt to load in patient folder previously loaded onto rosa brain account, got failed to load error.This only occurred on rosa brain account and not upon attempt to load in maintenance launch of rosa software.In the rosa logs, the error code read 'failed to read.Hdr file'.Surgery was cancelled due to other circumstance thus, did not cause a delay.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROSA ONE
Type of Device
COMPUTER-ASSISTED SURGICAL DEVICE
Manufacturer (Section D)
MEDTECH SA
zac eureka
900 rue du mas de verchant
montpellier, languedoc-roussillon 34000
FR  34000
Manufacturer (Section G)
MEDTECH SA
zac eureka
900 rue du mas de verchant
montpellier, languedoc-roussillon 34000
FR   34000
Manufacturer Contact
jay sharma
1520 tradeport drive
jacksonville, FL 32218
9047414400
MDR Report Key12179795
MDR Text Key261951981
Report Number3009185973-2021-00162
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K200511
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 07/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberROSA ONE
Device Catalogue NumberROSAS00203
Device Lot Number3.1.4.1650
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/23/2021
Initial Date FDA Received07/16/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/12/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
-
-