DATASCOPE CORP. - MAHWAH CS100 INTRA-AORTIC BALLOON PUMP, ENGLISH, 220V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-3013-55 |
Device Problem
Connection Problem (2900)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/25/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Testing of actual/suspected device (10/213): a getinge field service engineer (fse) was dispatched to investigate.The fse evaluated the iabp unit, but was unable to reproduce the reported issue.The fse found that the iabp was able to retrieve the ecg signal using an external signal cable normally.The fse then performed functional and safety checks to meet factory specifications.The iabp was then released to the customer and cleared for clinical service.A supplemental report will be submitted if additional information is provided.
|
|
Event Description
|
It was reported that prior to patient use, the cs100 intra-aortic balloon pump (iabp) external signal cable failed.There was no patient involvement, and no adverse event reported.
|
|
Manufacturer Narrative
|
Analysis of production: (3331) the device history record review concluded that there were no ncmrs, rework, or deviations documented for the reported lot/serial number.Based on the dhr/lhr review results, it was determined that there is no relation between the manufacturing process and the reported failure.Historical data analysis: (4109) the review of the historical data indicates that no other similar complaint was reported for the same lot/serial number and reported failure mode.Trend analysis: (4110) the overall 24 month product complaint trend data for the period jul-2019 through jun-2021 was reviewed.There were no triggers identified for the review period.
|
|
Search Alerts/Recalls
|
|
|