DATASCOPE CORP. - MAHWAH CS100 SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-3013-55
Device Problem Connection Problem (2900)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/25/2021
Event Type  malfunction  
Manufacturer Narrative
Testing of actual/suspected device (10/213): a getinge field service engineer (fse) was dispatched to investigate. The fse evaluated the iabp unit, but was unable to reproduce the reported issue. The fse found that the iabp was able to retrieve the ecg signal using an external signal cable normally. The fse then performed functional and safety checks to meet factory specifications. The iabp was then released to the customer and cleared for clinical service. A supplemental report will be submitted if additional information is provided.
 
Event Description
It was reported that prior to patient use, the cs100 intra-aortic balloon pump (iabp) external signal cable failed. There was no patient involvement, and no adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameCS100
Type of DeviceSYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
dorota wolpiuk
1300 macarthur blvd
mahwah, NJ 
MDR Report Key12180050
MDR Text Key261935425
Report Number2249723-2021-01543
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeTH
PMA/PMN Number
K031636
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 07/14/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received07/16/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number0998-00-3013-55
Device Catalogue Number0998-00-3013-55
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received06/25/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/01/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device?
Type of Device Usage Reuse

-
-