• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VERYAN MEDICAL LTD BIOMIMICS 3D VASCULAR STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VERYAN MEDICAL LTD BIOMIMICS 3D VASCULAR STENT SYSTEM Back to Search Results
Model Number 144700-05
Medical Device Problem Codes Difficult or Delayed Positioning (1157); Difficult or Delayed Activation (2577); Malposition of Device (2616); Positioning Problem (3009); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/16/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is related to mdr number: 3011632150-2021-00031.The device was not returned for evaluation.The complaint investigation is ongoing.When further information regarding this event becomes available, or the investigation is concluded a follow-up/supplemental report will be submitted.
 
Event or Problem Description
This mdr report is related to mdr number 3011632150-2021-00031.The procedure in question involved the implantation of two 150mm biomimics 3d stents to treat a long target lesion from the distal superficial femoral artery (sfa) to popliteal artery.Following stent delivery system (sds) preparation, the sds was tracked over the iliac bifurcation to the proximal popliteal artery using a 6fr x 45cm flexor ansel 1 curve ancillary sheath over a 0.014" guidewire.There was severe vessel tortuosity and a tight bifurcation angle.A cross-over approach was used, and it was reported that no resistance was felt during the approach of the device to the target site.Upon initiation of deployment, the physician experienced resistance and at first was unable to retract the outer braid.The sds was slightly retracted to remove slack and the tuohy borst was unlocked.The physician made several attempts to continue the deployment, and used significant force throughout the deployment procedure to get the stent to release.The physician also needed to pause and reposition the device at the target site between deployment attempts.The first stent implanted (a 5 x 150mm stent) was positioned slightly more proximal to the intended position.The second stent (6 x 150mm stent) was then deployed with a slight overlap proximally to the first stent and resistance was also reported but successful deployment was achieved after a small number of attempts.There is no reported patient impact and both stents remain implanted.
 
Additional Manufacturer Narrative
This mdr number is related to 3011632150-2021-00031.The device history review was completed.There were no issues identified that could have contributed to the reported complaint.The lot met the relevant in-process monitoring acceptance criteria.The coding has also been updated to reflect completion of this investigation.There were no devices returned for evaluation and no angiographic imaging returned despite multiple requests made by veryan's complaint investigators.The conclusions are based on the information provided in the initially reported complaint as well as descriptions provided by the veryan representative which included his suggestions to the implanting physician during the procedure and the results of the complaint investigation.Although the physician reported significant resistance during the deployment of both stents, the complaint investigation concluded that this is likely to be related to the severely tortuous patient anatomy and a tight bifurcation angle.These types of anatomical conditions can lead to higher deployment forces due to the likelihood of friction created between components of the biomimics 3d delivery system (e.G.Outer braid and inner assembly) as well as between the ancillary devices used with the biomimics 3d system (e.G.Guidewire/introducer sheath).The updated information still highlights that a number of attempts to complete deployment were required which included repositioning the device at the target site.The veryan sales representative confirmed that there was no distortion/misalignment/foreshortening of the first stent.It was implanted slightly proximal to what was intended.There is no indication of a product quality issue with respect to the design, manufacture or labelling of the device.If further information regarding this event becomes available, a follow-up report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOMIMICS 3D VASCULAR STENT SYSTEM
Common Device Name
BIOMIMICS 3D VASCULAR STENT SYSTEM
Manufacturer (Section D)
VERYAN MEDICAL LTD
unit 11,galway technology park
parkmore road
galway, H91 V E0H
EI  H91 VE0H
MDR Report Key12180823
Report Number3011632150-2021-00030
Device Sequence Number2696388
Product Code NIP
UDI-Device Identifier05391526850503
UDI-Public(01)05391526850503(17)220424(11)200923(10)0000052818
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P180003
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Expiration Date04/24/2022
Device Model Number144700-05
Device Catalogue Number144700-05
Device Lot Number0000052818
Initial Date Received by Manufacturer 06/16/2021
Supplement Date Received by Manufacturer08/11/2021
Initial Report FDA Received Date07/16/2021
Supplement Report FDA Received Date09/08/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-