• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION AMS 800 URINARY CONTROL SYSTEM; DEVICE INCONTINENCE MECHANICAL/HYDRAULIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION AMS 800 URINARY CONTROL SYSTEM; DEVICE INCONTINENCE MECHANICAL/HYDRAULIC Back to Search Results
Model Number 72400023
Medical Device Problem Code Material Puncture/Hole (1504)
Health Effect - Clinical Code Urinary Incontinence (4572)
Date of Event 06/10/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Investigation summary: based on the information available, the cause that contributed to the reported balloon tear cannot be established as the product is not available for analysis.Device history record (dhr) : the device history record (dhr) confirmed that the device met all material, assembly and performance specifications.Device technical analysis : the device is not available for analysis; therefore, no physical or visual analysis of the product could be performed.The reported device performance allegation cannot be confirmed.Labeling review: a review of the device instructions for use (ifu) was completed and did not reveal any evidence of device misuse, off label use, or failure to follow instructions.Investigation conclusion: based on the information available, a conclusion code of cause not established was assigned to this investigation.
 
Event or Problem Description
It was reported that the patient had a surgical procedure to revise an artificial urinary sphincter(aus) due to recurring incontinence two and a half years after the aus was implanted.When the aus was explanted, a hole was found in the prb.A new pump and prb was implanted and the device was tested and fully functioning after verification by cystoscope.No further patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMS 800 URINARY CONTROL SYSTEM
Common Device Name
DEVICE INCONTINENCE MECHANICAL/HYDRAULIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
10700 bren road w
minnetonka MN 55343
MDR Report Key12180962
Report Number2124215-2021-19788
Device Sequence Number909679
Product Code EZY
UDI-Device Identifier00878953000619
UDI-Public00878953000619
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P000053
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date05/30/2023
Device Model Number72400023
Device Catalogue Number72400023
Device Lot Number1000114184
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/22/2021
Initial Report FDA Received Date07/16/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age67 YR
-
-