Intuitive surgical, inc.(isi) received the small grasping retractor instrument involved with this complaint and completed the device evaluation.Failure analysis investigations confirmed the customer reported complaint.Failure analysis found the primary failure of broken pitch cable to be related to the customer reported complaint.The instrument was found to have a broken pitch cable at the distal end.The broken cable segment that contains the crimp was missing from the clevis.Pitch cable breakage occurs when tensile load exceeds the ultimate strength of the material.The pitch cable construction is designed to optimize load and fatigue (cycling) characteristics.Variation in customer use conditions, procedure type, patient anatomy, product handling, instrument tip lengths, grip torque, and manufacturing tolerances are a few variables which can influence pitch cable failure.In addition, the instrument was found to have indentations on the cable crimp.The cable crimp appeared to be flattened.The root cause was not established.A review of the site's complaint history does not reveal any related complaints involving this product or this event.A review of system logs for system sh0659 with a procedure date of (b)(6) 2021, confirmed a small grasping retractor (pn# 420318-07 / lot # n10210106 - 880) was used during this procedure.The instrument has maximum 10 uses.There were 9 more uses remaining.A review of the site's system logs for the reported procedure date was conducted by regulatory post market surveillance (rpms) analyst.Investigation revealed there were no related system errors to have occurred during the surgical procedure that would have likely caused or contributed to the reported complaint.The small grasping retractor is a multiple-use endoscopic instrument with a grasping tip to be used in conjunction with the da vinci system.The instrument is designed to grab, manipulate, retract, and dissect tissue during a da vinci assisted surgical procedure.Based on the additional information obtained from failure analysis investigations, this complaint is being reported due to the following conclusion: this instrument is designed with two pitch cables, each with a crimp at the distal end.If a pitch cable breaks at the distal end, a cable segment and/or the crimp could fall into the patient.While there was no harm or injury to the patient, the reported failure mode could likely cause or contribute to an adverse event if it were to recur.
|
|
It was reported that during a da vinci-assisted surgical procedure, the small grasping retractor instrument cable was broken.The procedure was completed by using a back-up instrument and there was no reported injury.Intuitive surgical, inc.(isi) made multiple follow-up attempts to obtain additional information.However, no further details have been received as of the date of this report.
|