• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTUITIVE SURGICAL, INC ENDOWRIST;DAVINCI SI; SMALL GRASPING RETRACTOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTUITIVE SURGICAL, INC ENDOWRIST;DAVINCI SI; SMALL GRASPING RETRACTOR Back to Search Results
Model Number 420318-07
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Intuitive surgical, inc.(isi) received the small grasping retractor instrument involved with this complaint and completed the device evaluation.Failure analysis investigation confirmed the customer reported complaint.The instrument was found to have a broken pitch cable at the distal end.The broken cable segment that contains the crimp was still installed in the clevis.Pitch cable breakage occurs when tensile load exceeds the ultimate strength of the material.The pitch cable construction is designed to optimize load and fatigue (cycling) characteristics.Variation in customer use conditions, procedure type, patient anatomy,product handling, instrument tip lengths, grip torque, and manufacturing tolerances are a few variables which can influence pitch cable failure.A review of the instrument log for the small grasping retractor instrument (part# 420318-07/n10201124 008) associated with this event has been performed.Per logs, the instrument was last used on 28-01-2021 on system sh0659, with 8 uses remaining.It is unknown if this is the date of the reported event or not as the event date was not provided by the customer.There was no photo or video provided by the site for review.A review of the site's complaint history does not show any additional complaints related to this product and/or this event.This complaint is being reported due to the following conclusion: this instrument is designed with two pitch cables, each with a crimp at the distal end.If a pitch cable breaks at the distal end, a cable segment and/or the crimp could fall inside the patient.Although there was no patient injury, the product malfunction could cause or contribute to an adverse event if the failure were to recur.
 
Event or Problem Description
It was reported that during a da vinci-assisted surgical procedure on an unknown event date, the small grasping retractor instrument cable was broken.A similar backup da vinci instrument was used to continue with the case.The procedure was completed with no reported injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOWRIST;DAVINCI SI
Common Device Name
SMALL GRASPING RETRACTOR
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
MDR Report Key12186946
Report Number2955842-2021-10813
Device Sequence Number16983520
Product Code NAY
UDI-Device Identifier00886874111802
UDI-Public(01)00886874111802(10)N10201124
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
K050369
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,other
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 06/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number420318-07
Device Catalogue Number420318
Device Lot NumberN10201124 008
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/16/2021
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/21/2021
Initial Report FDA Received Date07/19/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DA VINCI INSTRUMENTS AND ACCESSORIES
-
-