Model Number 3186
Device Problem Patient-Device Incompatibility (2682)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  Injury  
Manufacturer Narrative
Event date is unknown.
 
Event Description
Related manufacturer report number: 1627487-2021-15622.It was reported that the patient was experiencing ineffective stimulation due to impedance issues.It was noted that the patient had excessive scar tissue that surrounding the implanted leads.As result, the entire scs system was explanted.
 
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OCTRODE LEAD KIT, 60CM LENGTH
Type of Device
SCS LEAD
Manufacturer (Section D)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
MDR Report Key12187939
MDR Text Key262167130
Report Number1627487-2021-15623
Device Sequence Number1
Product Code LGW
UDI-Device Identifier05415067017246
UDI-Public05415067017246
Combination Product (y/n)N
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Type of Report Initial,Followup
Report Date 07/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2017
Device Model Number3186
Device Catalogue Number3186
Device Lot Number5018286
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/15/2021
Initial Date FDA Received07/19/2021
Supplement Dates Manufacturer Received07/15/2021
Supplement Dates FDA Received07/19/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
SCS LEAD, MODEL 3186; SCS LEAD, MODEL 3186
Patient Outcome(s) Other;
Patient Weight104
-
-