| Model Number |
03.010.441 |
| Medical Device Problem Code |
Device-Device Incompatibility (2919)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Additional narrative: patent identifier: mrn: (b)(6).The device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that on unknown date, the patient underwent for a procedure.During the procedure, while reaming a suprapatellar tibial nail, there was difficulty to catching the flexible reamers on the protection sleeve when retracting flexible reamer.Furthermore, when drilling for the s1 proximal interlocking screw, the drill bit would not pass through the nail and unable to place a screw through the desired hole.The s2 proximal interlocking screw was successfully drilled and inserted along with another proximal oblique screw.There was 10 minutes of surgical delay noted.No fragments generated noted.The surgery was successfully completed.The patient outcome was unknown.Concomitant devices reported: unk - insertion instruments: trauma :part# unknown, lot# unknown, qty unknown).Unk - reaming rods :part# unknown, lot# unknown, qty unknown).This complaint involves three (3) devices.This report is for (1) aiming arm for suprapatellar.This report is 1 of 5 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: this complaint is not confirmed as there was no damage or defect observed on the returned aiming arm.As other mating devices were not returned, a functional assessment could not be performed.No root cause could definitively be determined for the reported complaint condition.No new, unique or different patient harms were identified as a result of this evaluation.There was no indication that a design or manufacturing issue contributed to the complaint.No design issues were observed during the document/specification review.Based on the investigation findings, it has been determined that no corrective and/or preventive action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Device history lot : part: 03.010.441.Lot: 15-9957.Manufacturing site: selzach.Supplier: leitner ag.Release to warehouse date: 10 july 2015.A manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.Review of the device history record showed that there were no issues during the manufacture of this product, and any subcomponents, which would contribute to this complaint condition.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|
|