• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC FLEXOR ANSEL GUIDING SHEATH; DYB INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC FLEXOR ANSEL GUIDING SHEATH; DYB INTRODUCER, CATHETER Back to Search Results
Model Number G29982
Medical Device Problem Codes Material Separation (1562); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/30/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
 
Event or Problem Description
As reported, during a renal angiogram and pta procedures a flexor ansel guiding sheath was used.While pulling the sheath out to replace it with a short sheath, the sheath snapped.No unintended section of the device remained inside of the patient's body.No additional procedures were required due to the occurrence.No adverse effects were report due to the occurrence.Additional patient outcome and event details have been requested.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Additional/corrected information: b5, h6 (annex a).H3: device evaluated by mfg = other (81) - device evaluation has begun; however, a conclusion is not yet available.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
Upon return and initial evaluation of the complaint device, the entire sheath was noted to be separated from the hub of the device.
 
Event or Problem Description
No new patient or event information to report.
 
Additional Manufacturer Narrative
Description of event: as reported, during a renal angiogram and pta procedures a flexor ansel guiding sheath was used.While pulling the sheath out to replace it with a short sheath, the sheath separated from the hub of the device.Investigation ¿ evaluation.A visual inspection of the returned device was conducted.A document based investigation was also performed including a review of complaint history, device history record, the instructions for use, manufacturing instructions, and quality control data.The customer returned one kcfw-6.0-18/38-45-rb-anl1-hc to cook for investigation.Physical examination of the returned device showed: visual inspection results, one device returned in a used and damaged condition.Separation of entire catheter shaft, including flare was confirmed no material was within check flo valve.A review of the device history record found no non-conformances related to the reported failure mode.Because there are no related non-conformances, adequate inspection activities have been established, there is objective evidence that the dhr was fully executed, and no other lot related complaints that have been received from the field, it was concluded that there is no evidence that nonconforming product exists in house or in field.The instructions for use (ifu), provides the following information to the user related to the reported failure mode: "warnings¿ ¿if resistance is encountered during advancement of flexor sheath, assess cause of resistance and consider dilation of any restriction identified or consider alternate treatment strategy.If flexor sheath is advanced through resistance, force to remove the sheath will be higher, increasing the risk of sheath material or hub separation upon withdrawal.¿ ¿reinsertion of dilator prior to removal of flexor sheath increases the strength of the sheath and lessens the risk of device separation.If resistance i anticipated or encountered during withdrawal of flexor sheath, consider carefully reinserting the dilator prior to continuing removal.¿ ¿precautions¿ ¿in order to ensure device compatibility, choose a sheath size large enough to accommodate the maximum outer diameter of any devices that will be placed through the sheath.¿ ¿all interventional or diagnostic instruments used with this product should move freely through the valve and sheath to avoid damage.¿ ¿sheath removal¿ ¿insert a wire guide until its tip extends at least 10cm past the tip of the sheath.¿ ¿remove the sheath.Avoid applying traction to the hub during removal.If resistance is anticipated or encountered during withdrawal of the flexor sheath, consider reinserting the dilator and removing the sheath and dilator as a unit.¿ ¿remove the wire guide.¿ ¿how supplied¿ ¿upon removal from package, inspect the product to ensure no damage has occurred.¿ a review of relevant manufacturing documents was conducted.It was concluded that the device aspect in question was visually/functionally inspected by quality control and no related gaps in production or processing controls were noted.Cook has concluded a component failure contributed to this failure mode.Per the quality engineering risk assessment, no further action is warranted.Cook will continue to monitor this device via the complaints database for similar complaints.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXOR ANSEL GUIDING SHEATH
Common Device Name
DYB INTRODUCER, CATHETER
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key12189131
Report Number1820334-2021-01830
Device Sequence Number5127862
Product Code DYB
UDI-Device Identifier00827002299825
UDI-Public(01)00827002299825(17)231015(10)13492024
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K142829
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/15/2023
Device Model NumberG29982
Device Catalogue NumberKCFW-6.0-18/38-45-RB-ANL1-HC
Device Lot Number13492024
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/26/2021
Initial Date Received by Manufacturer 07/06/2021
Supplement Date Received by Manufacturer07/26/2021
10/08/2021
Initial Report FDA Received Date07/19/2021
Supplement Report FDA Received Date08/16/2021
10/26/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age72 YR
-
-