MEDTRONIC PUERTO RICO OPERATIONS CO. SURESCAN; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR
|
Back to Search Results |
|
Model Number 97714 |
Device Problems
Energy Output Problem (1431); Improper or Incorrect Procedure or Method (2017); Battery Problem (2885); Impedance Problem (2950)
|
Patient Problems
Inadequate Pain Relief (2388); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/22/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
Concomitant medical products: product id: 977c165, lot#: none, serial# (b)(4), implanted: (b)(6) 2016, explanted: none, product type: lead.Other relevant device(s) are: product id: 977c165, serial/lot #: (b)(4), ubd: 21-apr-2020, udi#: (b)(4).If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Information was received from a patient via a manufacturer representative (rep) regarding a patient who was implanted with an implantable neurostimulator (ins).It was reported that the ins battery was depleted.The patient stated stimulation wasn¿t working.The patient didn¿t charge or use for over 6 months.It was determined that the battery was overdischarged.A physician mode recharge was performed and the power on reset (por) was cleared.The issue resolved.No symptoms were reported.Additional information received.Ins replaced (b)(6) 2021.Impedance issues on all 16 electrodes.Attempts were made to reprogram lead.Unsuccessful.Dr will refer to spinal surgeon for lead replacement.
|
|
Manufacturer Narrative
|
H10.Additional information regarding the impedance issues will be submitted as a supplemental under manufacturer report # 2649622-2 021-05639.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|