MAUDE Adverse Event Report: ABBOTT VASCULAR MINI TREK CORONARY DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR MINI TREK CORONARY DILATATION CATHETER Back to Search Results
Model Number 1012270-20
Device Problems Difficult to Remove (1528); Material Rupture (1546); Improper or Incorrect Procedure or Method (2017)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/22/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The customer reported the device is not returning.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event Description
It was reported that the device may not have been prepared prior to use.The procedure was to treat a lesion in the right coronary artery (rca).A 2x20mm mini trek balloon dilatation catheter (bdc) was advanced to the lesion without issue and inflated 1 time to 8 atmospheres (atm), when the balloon ruptured.Resistance was met while removing the bdc from the anatomy.A non-abbott device was used to successfully complete the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
 
Manufacturer Narrative
The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history of the reported lot did not indicate a lot specific quality issue.It was reported by the account that the balloon dilatation catheter was not soaked prior to use.It should be noted that the coronary dilatation catheters (cdc), trek rx and mini trek rx, global instructions for use states: submerge the balloon in sterile heparinized normal saline during balloon preparation, to activate the coating.In this case, it is unknown if the ifu violation caused or contributed to the reported complaint.The investigation was unable to determine a conclusive cause for the reported balloon rupture; however, the reported difficulty removing the device appears to be related to operational context.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MINI TREK CORONARY DILATATION CATHETER
Type of Device
CORONARY DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key12193170
MDR Text Key262467527
Report Number2024168-2021-06194
Device Sequence Number1
Product Code LOX
UDI-Device Identifier08717648138164
UDI-Public08717648138164
Combination Product (y/n)N
PMA/PMN Number
K103110
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 08/06/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2023
Device Model Number1012270-20
Device Catalogue Number1012270-20
Device Lot Number00805G1
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 06/26/2021
Initial Date FDA Received07/19/2021
Supplement Dates Manufacturer Received07/21/2021
Supplement Dates FDA Received08/06/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-