• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN KANGAROO EPUMP - NEW; PUMP, INFUSION, ENTERAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN KANGAROO EPUMP - NEW; PUMP, INFUSION, ENTERAL Back to Search Results
Model Number 382400
Medical Device Problem Code Excess Flow or Over-Infusion (1311)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/24/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The incident sample has been requested but to date has not been received for evaluation.  if the sample is received, or if additional information pertinent to the incident is obtained a follow-up report will be submitted.  as part of our manufacturing process, all device history records are reviewed and approved by quality, prior to release of product.
 
Event or Problem Description
The customer reported that the device infuses too fast.The volume to be infused (vtbi) outside of spec.Additional information was received stating that the exact amount is unknown, it was at least 5 ml overfed.The volume to be infused was set to 30 ml and the rate was set to 60 ml/hr.The actual volume delivered was greater than 30 ml and the feed was completed in.5 hours.
 
Additional Manufacturer Narrative
The service history record was reviewed and indicated that this is the first time that this unit has been returned for service.An evaluation of the kangaroo pump was performed.The unit was triaged, and the reported issue was confirmed.After review, it was determined that the us sensor was damaged (based on error code index list: ultrasound voltage when tube is full in a/d counts).The sensor was replaced, and the problem was resolved.Also, battery was damaged due to obsolete, rotor damaged due to rollers not spinning freely.Gear box was damaged due to worn encoder 2, front housing damaged, back housing damaged due to screw boss area broken.Overlay, badge was replaced due to front housing replacement.Bump, hinge label, vynil foot and sn label were replaced due to back housing replacement.Power connection label replaced due to opening of the unit.No further actions are required at this time.Complaint trending information is being reviewed on a monthly basis and if a trend is observed, actions will be taken as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KANGAROO EPUMP - NEW
Common Device Name
PUMP, INFUSION, ENTERAL
Manufacturer (Section D)
COVIDIEN
15 hampshire street
mansfield MA 02048
MDR Report Key12193604
Report Number1282497-2021-10498
Device Sequence Number20045544
Product Code LZH
UDI-Device Identifier10884521006614
UDI-Public10884521006614
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model Number382400
Device Catalogue Number382400
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 06/24/2021
Supplement Date Received by Manufacturer06/24/2021
Initial Report FDA Received Date07/19/2021
Supplement Report FDA Received Date10/19/2021
Patient Sequence Number1
-
-