Model Number 10605
Medical Device Problem Codes Failure to Advance (2524); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/06/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that stent damage occurred.The 95% stenosed target lesion was located in the severely tortuous and moderately calcified proximal circumflex artery.Following pre-dilatation with a semi-compliant and non-compliant balloons, a 3.00 x 28 synergy drug-eluting stent was advanced but failed to cross the lesion.The physician used a non-boston scientific stent which also failed to cross.Then, a 3.00 x 16 synergy xd drug-eluting stent was advanced but failed to cross the lesion as well.A 1.5 mm rotapro burr was used and further dilatation was performed with a 3.00 non-compliant balloon catheter.Another 3.00 x 28 was advanced but would not cross the lesion.The device was removed and it was noticed that a stent strut was lifted.The same non-boston scientific stent was used to complete the procedure.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY
Common Device Name
BIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key12193669
Report Number2134265-2021-09244
Device Sequence Number19783859
Product Code NIQ
UDI-Device Identifier08714729840299
UDI-Public08714729840299
Combination Product (Y/N)Y
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
P150003
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/16/2022
Device Model Number10605
Device Catalogue Number10605
Device Lot Number0026404846
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/06/2021
Initial Report FDA Received Date07/19/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-