• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESMED LTD ASTRAL EXTERNAL BATTERY, ROW

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESMED LTD ASTRAL EXTERNAL BATTERY, ROW Back to Search Results
Model Number 27918
Device Problem Premature Discharge of Battery (1057)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/22/2021
Event Type  malfunction  
Manufacturer Narrative
The device was returned to resmed and an evaluation confirmed the complaint.The external battery was replaced to address this issue.The device was serviced and fully tested before it was returned to the customer.Based on all available evidence and complaint investigations of a similar nature, an investigation determined that the reported complaint was due to an isolated component failure within the device battery assembly.Resmed's risk analysis for this failure mode concludes that the risk is acceptable.Resmed reference #: (b)(4).
 
Event Description
It was reported to resmed that an astral external battery had a reduced level of capacity.There was no patient harm or serious injury reported as a result of this incident.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASTRAL EXTERNAL BATTERY, ROW
Manufacturer (Section D)
RESMED LTD
1 elizabeth macarthur drive
bella vista
sydney, nsw 2153
AS  2153
Manufacturer (Section G)
RESMED LTD
1 elizabeth macarthur drive
bella vista
sydney, nsw 2153
AS   2153
Manufacturer Contact
michelle hughes
1 elizabeth macarthur drive
bella vista
sydney, nsw 2153
AS   2153
MDR Report Key12194009
MDR Text Key264477460
Report Number3004604967-2021-00800
Device Sequence Number1
Product Code CBK
UDI-Device Identifier00619498279180
UDI-Public(01)00619498279180(10)1446246
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K152068
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 07/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/19/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number27918
Device Catalogue Number27918
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/24/2021
Is the Reporter a Health Professional? Yes
Distributor Facility Aware Date06/20/2021
Date Manufacturer Received06/20/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/08/2020
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-