MAUDE Adverse Event Report: BECTON DICKINSON MEDICAL SYSTEMS SYRINGE 10ML LL W/NDL 20X1 RB; PISTON SYRINGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON MEDICAL SYSTEMS SYRINGE 10ML LL W/NDL 20X1 RB; PISTON SYRINGE Back to Search Results
Catalog Number 309644
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code Needle Stick/Puncture (2462)
Date of Event 06/16/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation summary: two photos and two loose 10ml syringes with attached needles (p/n 309644) were received.The samples were visually evaluated.Both sample's hubs were twisted nearly to the point of snapping.Each hub was twisted in the same location.The observed condition was non-conforming per product specification.Batch #0265079 is considered in compliance with our product specification requirements.Furthermore, a device history record review showed no rejected inspections or quality issues during the production of the provided lot number that could have contributed to the reported defect.Investigation conclusion: complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly.Our business team regularly reviews the collected data for identification of emerging trends.Root cause description: potential root cause for the twisted hub defect is associated with the assembly process.These conditions are occurring at/below their expected frequency.Therefore, no corrective action is required at this time.Rationale: capa not required at this time.
 
Event or Problem Description
It was reported that syringe 10ml ll w/ndl 20x1 rb needle was damaged.The following information was provided by the initial reporter: it was reported that needle bent and stuck staff member while attempting to draw up medication.Customer response: did the staff member require any medical intervention? no.What was the date of the event? on (b)(6) 2021.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYRINGE 10ML LL W/NDL 20X1 RB
Common Device Name
PISTON SYRINGE
Manufacturer (Section D)
BECTON DICKINSON MEDICAL SYSTEMS
route 7 and grace way
canaan CT 06018
MDR Report Key12194014
Report Number1213809-2021-00497
Device Sequence Number3802416
Product Code FMF
UDI-Device Identifier00382903096442
UDI-Public00382903096442
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
K980987
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source other,user facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 07/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date08/31/2025
Device Catalogue Number309644
Device Lot Number0265079
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/28/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/21/2021
Initial Report FDA Received Date07/19/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-