| Catalog Number |
309644 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
Needle Stick/Puncture (2462)
|
| Date of Event |
06/16/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation summary: two photos and two loose 10ml syringes with attached needles (p/n 309644) were received.The samples were visually evaluated.Both sample's hubs were twisted nearly to the point of snapping.Each hub was twisted in the same location.The observed condition was non-conforming per product specification.Batch #0265079 is considered in compliance with our product specification requirements.Furthermore, a device history record review showed no rejected inspections or quality issues during the production of the provided lot number that could have contributed to the reported defect.Investigation conclusion: complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly.Our business team regularly reviews the collected data for identification of emerging trends.Root cause description: potential root cause for the twisted hub defect is associated with the assembly process.These conditions are occurring at/below their expected frequency.Therefore, no corrective action is required at this time.Rationale: capa not required at this time.
|
| |
|
Event or Problem Description
|
|
It was reported that syringe 10ml ll w/ndl 20x1 rb needle was damaged.The following information was provided by the initial reporter: it was reported that needle bent and stuck staff member while attempting to draw up medication.Customer response: did the staff member require any medical intervention? no.What was the date of the event? on (b)(6) 2021.
|
| |
|
Search Alerts/Recalls
|