MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATTUNE FEM POR PS RT SZ 6; ATTUNE CEMENTLESS IMPLANTS : KNEE FEMORAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATTUNE FEM POR PS RT SZ 6; ATTUNE CEMENTLESS IMPLANTS : KNEE FEMORAL Back to Search Results
Model Number 1504-11-206
Medical Device Problem Code Difficult to Insert (1316)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/28/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that surgeon complains about femur being very hard to seat/implant on patient.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
Additional information received stated that no instrument was broken, no loosening, and no adverse consequences that affect the patient.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot: product code 150411206, work order (b)(4) was manufactured on 16-feb-2021.11 parts were manufactured per specification and all raw materials met specification.Scrap: 1 part from this lot was scrapped: scrap code a016: base metal visible.This concerns base metal being visible through the porous coating on the parts.There is no correlation between this scrap reason and the failure mode of the complaint.Reprocessing: there was no material reprocessing reports (mrr) associated with this lot.Non-conformance: 1 non-conformances associated with this lot: nr-0158788 is related to a furnace burn that had been processed with a quantity of parts exceeding the maximum allowable quantity.Based on the analysis from metallurgy, r&d and fatigue testing that was carried out it was determined that there was no impact to product design or safety and no impact to patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE FEM POR PS RT SZ 6
Common Device Name
ATTUNE CEMENTLESS IMPLANTS : KNEE FEMORAL
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
MDR Report Key12194081
Report Number1818910-2021-15474
Device Sequence Number3397700
Product Code MBH
UDI-Device Identifier10603295042051
UDI-Public10603295042051
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P830055
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1504-11-206
Device Catalogue Number150411206
Device Lot Number9698752
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/28/2021
Supplement Date Received by Manufacturer07/26/2021
08/26/2021
Initial Report FDA Received Date07/19/2021
Supplement Report FDA Received Date07/29/2021
08/26/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-