MMT-751LNAS |
| Medical Device Problem Codes |
Break (1069); Battery Problem (2885); Moisture or Humidity Problem (2986); Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
02/26/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Insulin pump monitored for several days.Unit received with all operating currents within spec and passed displacement test.No unexpected low battery alarm anomalies noted.However, moisture damage at electronic assembly.Also, moisture damage motor during visual inspection.Pump received with minor scratched lcd window and cracked reservoir tube lip.The test infusion set and reservoir does lock into place.(b)(4).
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that insulin pump had condensation inside the screen.No harm requiring medical intervention was reported.The insulin pup had bad readings at times and frequency of the battery was changed.The device will not be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
The initial report was submitted with the wrong code in section h6.The correction has been made in this report.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
| |
|