| Model Number |
MMT-751NAL |
| Medical Device Problem Codes |
Break (1069); Computer Software Problem (1112); Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/28/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Insulin pump received with blank display due to corroded battery tube spring contact and corroded battery tube.Unable to perform all functional testing including the operating currents, a21 error test, rewind, basic occlusion, occlusion, prime/a33 and excessive no delivery test or verify unexpected error message due to blank display.Pump received with cracked battery tube threads, cracked reservoir tube lip and cracked case display window corner.The p-cap locks into the reservoir compartment properly noted.(b)(4).
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicates the insulin pump had a crack on the battery compartment and had many battery alarms.No harm to the customer requiring medical intervention reported.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
The initial report was submitted with the wrong code in section h6.The correction has been made in this report.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
| |
|